Haemodynamic monitoring

Arterial line waveform and damping

Reviewed Nov 2026·3 references
New· views rounded down, updated weekly

Set up, zero and level an arterial line, read the waveform, and use the fast-flush (square-wave) test to detect over- and under-damping.

Definition
An arterial line transmits the pressure wave through fluid-filled tubing to a transducer. Accuracy depends on correct zeroing, levelling at the phlebostatic axis and an adequately damped system. MAP is the most robust value even when damping is imperfect.

At the bedside

  1. 1

    Indications

    Shock on vasopressors, frequent ABGs, labile BP (e.g. hypertensive emergency, dissection), raised ICP, major surgery. Usually radial (check ulnar collateral by ultrasound/clinically), then femoral or brachial.

  2. 2

    Level and zero

    Transducer at the phlebostatic axis (4th intercostal space, mid-axillary line) with patient supine; for neuro patients who need cerebral perfusion pressure, some units level at the tragus (be consistent). Open to air and zero at set-up, after moving, and at least once per shift.

  3. 3

    Check the system

    Pressure bag inflated to 300 mmHg (flush ~3 mL/h), no air bubbles or clots, short non-compliant tubing, no kinks or three-way taps with stagnant blood.

  4. 4

    Do the fast-flush (square-wave) test

    Pull the flush tab for 1 s and release. Optimal: 1-2 oscillations before returning to baseline. Over-damped: < 1 oscillation, slurred upstroke → falsely low SBP, high DBP. Under-damped: > 2 oscillations, spiky trace → falsely high SBP, low DBP.

  5. 5

    Read the waveform

    Sharp upstroke (contractility), dicrotic notch (aortic valve closure), area under curve (stroke volume). Narrow pulse pressure suggests low SV or tamponade; wide pulse pressure suggests vasodilation or AR; respiratory variation suggests preload responsiveness.

  6. 6

    Troubleshoot

    Flat/dampened trace: check patient first (pulse, BP cuff), then flush, aspirate, reposition wrist, check bubbles, re-zero. Compare with NIBP if in doubt.

  7. 7

    Safety

    Label the line clearly as arterial (never inject drugs), keep it visible, check distal perfusion hourly, remove when no longer needed.

Square-wave test interpretation

PatternOscillationsEffect on readingsCauses
Optimal1-2Accurate-
Over-damped< 1, slow returnSBP falsely low, DBP falsely high, MAP OKAir bubbles, clot, kink, low bag pressure, compliant tubing
Under-damped> 2, ringingSBP falsely high, DBP falsely low, MAP OKLong tubing, too many connectors, tachycardia, stiff arteries
Red flag · call your senior now
Pale, cold or mottled hand distal to a radial line: remove the cannula immediately and assess for arterial thrombosis.
Red flag
Never inject drugs into an arterial line: risk of limb ischaemia and gangrene.
Pitfall
Transducer too low reads falsely high (~7.5 mmHg per 10 cm), too high reads falsely low.
Pitfall
Titrating vasopressors to a systolic pressure on a poorly damped trace. Use MAP.
Pearl
A sudden drop in arterial pressure with a normal NIBP and a flat trace is usually a technical problem; a sudden drop with loss of pulse is cardiac arrest.