Perioperative and cardiac surgical ICU

Epicardial pacing wires and temporary pacing

Reviewed Nov 2026·2 references
New· views rounded down, updated weekly

How to read and set a temporary pacing box after cardiac surgery: modes (AAI, VVI, DDD, asynchronous), output and sensitivity thresholds, daily checks, troubleshooting failure to capture or sense, and safe wire removal.

Definition
Temporary epicardial wires are stitched to the atrial and/or ventricular epicardium at the end of cardiac surgery to treat bradyarrhythmias, optimise cardiac output (AV synchrony, rate) and terminate some tachyarrhythmias. Their thresholds rise over days because of inflammation and fibrosis, and many fail by day 4–5.

At the bedside

Timed steps
  1. 1

    Identify the wires

    Arrival

    Confirm which wires are atrial and which ventricular (by convention atrial often exit right of the sternum and ventricular left — but always check the operation note/labels). Connect to the pacing box with labelled cables; keep unused wires insulated and dry.

  2. 2

    Record the settings

    Arrival

    Document mode, rate, atrial and ventricular output (mA), sensitivity (mV), AV delay and underlying rhythm on the chart. Check the box battery and have a spare battery and transcutaneous pacing pads available.

  3. 3

    Select the mode

    Choose

    Sinus bradycardia with intact AV conduction → AAI. AF/flutter with slow ventricular response → VVI. Heart block with atrial activity → DDD. Asynchronous (AOO/VOO/DOO) only temporarily (e.g. diathermy interference) — risk of R-on-T VF.

  4. 4

    Threshold testing

    Daily

    Capture threshold: at a rate above intrinsic, reduce output until capture is lost; set output at least 2–3 × threshold. Sensing threshold: set rate below intrinsic, increase the mV number until sensing fails (pacing spikes appear); set sensitivity at about half that value. Do NOT test capture in pacing-dependent patients without senior help.

  5. 5

    Failure to capture

    Problem

    Spikes without P/QRS: increase output, check connections and battery, reverse polarity, correct K⁺/acidosis/ischaemia; if wires failing, use a skin needle as an indifferent electrode (unipolar), start transcutaneous pacing or isoprenaline, arrange transvenous pacing.

  6. 6

    Failure to sense / oversensing

    Problem

    Undersensing (inappropriate spikes on intrinsic beats; risk of R-on-T): make more sensitive (lower mV). Oversensing (pauses because noise/T-waves inhibit pacing; crosstalk in DDD): make less sensitive, check leads, shorten ventricular blanking per expert advice.

  7. 7

    Overdrive pacing (expert)

    Atrial flutter

    Type 1 atrial flutter post-op may be terminated by rapid atrial pacing via atrial wires (e.g. 10–20% above flutter rate) — only with ventricular wire backup and senior/cardiology supervision; never deliver rapid pacing through ventricular wires.

  8. 8

    Removal

    Day 3–5

    Remove when no longer needed (often day 3–5, before discharge). Check INR, platelets and that heparin/LMWH is timed appropriately; pull gently with steady traction (cut flush at skin if resistant, per surgeon). Bed rest and monitor BP/HR for ~1 h; watch for tamponade.

Bradycardia while pacing is being set up

Atropine

IV
Dose
0.5–1 mg; repeat every 3–5 min to max 3 mg
Worked example
0.6 mg ampoule IV, repeat to 3 mg.

Bradycardia with failing wires — bridge to transvenous pacing

Isoprenaline (isoproterenol)

IV infusion
Dose
0.5–10 mcg/min
Range
0.5–10 mcg/min
Preparation
2 mg in 50 mL = 40 mcg/mL (check ampoule strength)
Worked example
2 mcg/min = 120 mcg/h = 3 mL/h of 40 mcg/mL.
Open infusion calculator

Pacing code (NBG) — first three letters

PositionMeaningLetters
1stChamber pacedA = atrium, V = ventricle, D = dual, O = none
2ndChamber sensedA, V, D, O
3rdResponse to sensingI = inhibited, T = triggered, D = dual (inhibit + trigger), O = none

Examples: AAI = paces and senses atrium, inhibited by native P waves. VVI = paces/senses ventricle. DDD = paces and senses both, tracks atrial activity. DOO = asynchronous dual-chamber.

Typical starting settings after cardiac surgery (guide)

SettingAtrialVentricular
Rate80–90/min (AAI/DDD)80–90/min (VVI backup often 50–60 if intrinsic rhythm adequate)
Output≥2–3 × capture threshold (often 5–10 mA initially)≥2–3 × capture threshold (often 5–10 mA initially)
SensitivityAbout half the sensing threshold (often 0.5–1 mV)About half the sensing threshold (often 2–3 mV)
AV delay (DDD)Commonly 150–180 ms (optimise to cardiac output)—

Settings are institution-specific; record the measured thresholds daily.

Troubleshooting

ProblemECG appearanceAction
Failure to captureSpikes without P or QRS↑ output, check connections/battery, reverse polarity, correct K⁺, unipolar via skin needle, transcutaneous pacing
Failure to pace (no output)No spikes, bradycardiaCheck battery/box/cables; oversensing? reduce sensitivity; asynchronous mode if needed
UndersensingSpikes on top of intrinsic beats/T wavesLower the mV value (more sensitive); consider turning pacing off if intrinsic rhythm adequate
Oversensing / crosstalkPauses, irregular pacingRaise the mV value (less sensitive), check for electrical interference
Pacemaker-mediated tachycardia (DDD)Fast paced rhythm at upper rateChange to VVI/DVI or adjust PVARP per expert
Red flag · call your senior now
Pacing-dependent patient (no underlying rhythm): never switch off or disconnect for testing without transcutaneous pacing ready.
Red flag
Spikes falling on T-waves (undersensing) can trigger VF — fix sensitivity or switch off pacing if the intrinsic rhythm is adequate.
Red flag
Hypotension and tachycardia after wire removal = tamponade until proven otherwise: echo and call the surgeon.
Pitfall
Handling exposed wire tips with ungloved hands or leaving them in contact with metal: microshocks can induce VF.
Pitfall
Forgetting that a pacing box set to asynchronous mode during surgery is still in DOO on ICU arrival — check and change to a demand mode.
Pitfall
Removing wires while the patient is fully anticoagulated (high INR or just after LMWH dose).
Pearl
AV-sequential pacing (DDD/DVI) can improve cardiac output by 15–20% compared with VVI in stiff post-operative ventricles that depend on the atrial kick.
Pearl
Atrial electrogram (connect atrial wires to a precordial ECG lead) helps diagnose wide- or narrow-complex tachycardia after surgery.

Calculators for this topic

Test yourself

Module quiz · score 70% to complete the module

Educational reference for trained clinicians. Check doses against your hospital protocol and the patient in front of you.